21 CFR Part 803 is the FDA’s regulation on Medical Device Reporting (MDR). This regulation outlines the requirements for manufacturers, importers, and device user facilities to report certain adverse events and product problems involving medical devices to the FDA. 

1. Scope and Purpose of 21 CFR Part 803

  • This regulation applies to:
    • Manufacturers of medical devices.
    • Importers of medical devices.
    • Device user facilities, including hospitals, nursing homes, and outpatient facilities.

The purpose of 21 CFR Part 803 is to ensure that the FDA is made aware of serious device-related adverse events and product malfunctions so that appropriate safety actions can be taken, including product recalls, safety alerts, or changes to device labeling.

2. Key Definitions in 21 CFR Part 803

Medical Device: Any instrument, apparatus, implement, machine, contrivance, implant, in vitro reagent, or other similar or related article, including any component, part, or accessory, which is intended for diagnosis, treatment, or prevention of disease or other conditions in humans or animals.

Adverse Event: Any incident in which a device may have caused or contributed to a death or serious injury.

Serious Injury: An injury or illness that:

  • Is life-threatening.
  • Results in permanent impairment or damage to a body function or structure.
  • Requires medical or surgical intervention to prevent permanent damage or impairment.

Device Malfunction: A failure of a device to meet its performance specifications or otherwise perform as intended. A malfunction that could potentially lead to death or serious injury if it were to recur must also be reported.

3. Reporting Requirements Under 21 CFR Part 803

There are different reporting requirements depending on whether the party involved is a manufacturer, importer, or user facility. Each party has specific responsibilities when it comes to reporting device-related adverse events and malfunctions.

A. Manufacturers

Manufacturers have the most comprehensive reporting obligations under 21 CFR Part 803.

Reportable Events:

  • Death or Serious Injury: If a medical device is suspected to have caused or contributed to a death or serious injury, the manufacturer must submit a report to the FDA.
  • Device Malfunction: If a device malfunctions and the malfunction would likely cause or contribute to a death or serious injury if it were to recur, it must also be reported.

Timeline for Reporting:

  • 30-Day Reports: Manufacturers must submit a Medical Device Report (MDR) within 30 calendar days of becoming aware of an event.
  • 5-Day Reports: If an event requires remedial action to prevent unreasonable risk of substantial harm to public health, or if the FDA requests the report, it must be submitted within 5 calendar days.

Follow-Up Reports:

  • Manufacturers must submit follow-up reports if they obtain additional information that was not available at the time of the initial report.

Reporting Format:

  • Manufacturers must submit MDRs electronically using FDA’s Electronic Submissions Gateway (ESG), following specific formats like the FDA Form 3500A.

B. Importers

Importers of medical devices also have reporting obligations under 21 CFR Part 803.

Reportable Events:

  • If an importer becomes aware that a device they imported caused or contributed to a death or serious injury, they must report this to the FDA.
  • If a device malfunctions and would likely cause or contribute to a death or serious injury if it were to recur, the importer must report this to the manufacturer.

Timeline for Reporting:

  • Reports of death or serious injury must be submitted to the FDA and the manufacturer within 30 calendar days of becoming aware of the event.
  • Malfunction reports must be submitted to the manufacturer within 30 calendar days.

C. Device User Facilities

User facilities such as hospitals, nursing homes, and outpatient clinics are also required to report certain adverse events under 21 CFR Part 803.

Reportable Events:

  • If a device user facility becomes aware of an incident where a device may have caused or contributed to a death, it must report this to both the FDA and the manufacturer.
  • If the device caused or contributed to a serious injury, the facility must report the event to the manufacturer. If the manufacturer is unknown, the event must be reported to the FDA.

Timeline for Reporting:

  • For death reports, the facility must report to the FDA and manufacturer within 10 working days of becoming aware of the event.
  • For serious injury reports, the facility must report to the manufacturer (or FDA, if unknown) within 10 working days.

Annual Summary Reporting:

  • User facilities are required to submit an annual report to the FDA summarizing all reportable events from the previous year. This report must be submitted by January 1st of each year, using FDA Form 3419.

4. Recordkeeping Requirements

All entities subject to 21 CFR Part 803 must maintain records of adverse events and associated documentation for a minimum of two years from the date of the event or for the lifetime of the device (if it is less than two years).

These records must include:

  • A copy of each report submitted to the FDA.
  • Information on each adverse event, including any investigations or corrective actions taken.

The FDA has the authority to inspect these records to ensure compliance with reporting requirements.

5. Exemptions and Variations

Certain devices and entities may be exempt from or subject to different reporting requirements under 21 CFR Part 803. For instance, devices under IDE (Investigational Device Exemption) or products covered under other regulations may have modified reporting standards.

Additionally, small manufacturers and importers can request a variance from the electronic reporting requirement, although they are still required to submit MDRs.

6. Compliance and Enforcement

Failure to comply with 21 CFR Part 803 can result in serious consequences, including:

  • Warning Letters from the FDA.
  • Product seizures or recalls.
  • Fines or penalties for failing to report adverse events or malfunctioning devices.
  • Criminal charges if non-compliance leads to significant patient harm.

The FDA actively monitors adverse event reports for trends or signals of new or increased risks associated with a device. If the FDA identifies a safety signal, it may require additional post-market surveillance, labeling changes, or even mandate a product recall.

7. FDA’s Role in Medical Device Reporting

The FDA uses the information collected under 21 CFR Part 803 to:

  • Monitor the safety and effectiveness of medical devices.
  • Detect and investigate potential device failures or emerging risks.
  • Take regulatory action when needed, such as issuing safety communications, recalls, or requesting additional data from manufacturers.

All submitted MDRs are stored in the FDA’s Medical Device Reporting (MDR) database, which is publicly accessible and allows healthcare professionals, patients, and researchers to review adverse events reported for specific devices.

8. Key Tools and Resources for Compliance

  • FDA Electronic Submissions Gateway (ESG): Used for submitting electronic MDRs.
  • FDA Form 3500A: Used for submitting adverse event reports for medical devices.
  • FDA Form 3419: Used for user facility annual reporting.

Manufacturers, importers, and user facilities are encouraged to implement strong internal systems for capturing, investigating, and reporting adverse events to ensure compliance with the FDA's reporting requirements.

Conclusion

21 CFR Part 803 is a critical regulation for ensuring the post-market safety of medical devices. It places the responsibility on manufacturers, importers, and device user facilities to report certain adverse events and malfunctions. Compliance with this regulation helps the FDA monitor device safety, detect emerging risks, and take corrective action when necessary, thereby protecting patient health and ensuring the continued safety of medical devices on the market.

Logo

© Copyright. All rights reserved.

Legal Notice| Privacy Policy

We need your consent to load the translations

We use a third-party service to translate the website content that may collect data about your activity. Please review the details in the privacy policy and accept the service to view the translations.